LogTag Ultra-Low Freezer Temp Data Logger

The LogTag® Ultra-Low Freezer Temp Data Logger is a great tool for monitoring & recording Ultra-Low-Temperatures in Freezers for blood plasma, vaccines, pharmaceutical and more over a range of -90°C to +40°C (-130°F to +104°F) via external probe. Excellent for tracking temperatures of COVID-19 vaccines, like those from Pfizer and BioNTech, which require storage and transport at temperatures of -70°C (-94°F).

Up to 15,900 temperature records—enough memory for 66 days at 6-minute logging—can be stored in the data recorder. If there have been any temperature deviations in the last 29 days or during the current day, the display is intended to indicate "At A Glance." In addition, the display provides the time, battery life, recording status, and current temperature measurement.

If the temperature readings exceed the pre-established alert levels, the alarm is set off. The audio alert will sound if it is setup, and the status indication will change from a check to an X. The day marker will also appear if it isn't already. The day marker stays in place to guarantee that a visible record of excursions is displayed on the display, even though the alarm status indicator and alarm can be easily removed. There is an option for a simplified alarm clearance, stop, and start routine.

A standard LogTag® Interface cradle can be used to download the logged data to LogTag® Analyzer operating software, which allows for an in-depth analysis of the excursions within the last 30 days. Alternatively, the statistics history displayed on the recorder's display can be examined directly.

Key Features

  • Built-in audible alarm
  • New, easy-to-read LCD display with 30-day summary in calendar format
  • Up to 15,900 recording – Enough memory for 66 days at 6-minute logging
  • Record inspections at the push of a button
  • Auto-pause function – unplugging the external probe automatically suspends alarm & statistics processing
  • External Probe with high-quality gold-plated connector
  • Comprehensive customization options including alarm settings, sample interval and trip duration
  • NIST Traceable Certificate


Specifications for LogTag Ultra-Low Freezer Temp Data Logger
External Sensor Measurement Range-90°C to +40°C (-130°F to +104°F).
Operating Temperature Range-30°C to +60°C (-22°F to +140°F).
Storage Temperature Range-10°C to +40°C (14°F to +104°F).
Recording Capacity15,905 temperature readings. 66 days @ 6min logging.
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics.
Sampling IntervalConfigurable from 30 seconds to hourly.
Logging Start OptionsPush button start or specific date & time. Optional start delay of up to 18 hours.
Recording IndicationState indicator “REC”.
Download TimeTypically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used.
EnvironmentalIP61 (when vertically mounted or hung).
Power Source3V;CR2032 (LiMnO2) coin cell battery, user replaceable.
Battery Life1 year storage, followed by 1 year of operation with normal use
Real Time ClockBuilt-in real time clock.
Standard Remote Sensor Cable LengthStandard: 1.5m (4’11″).
Extended: 3m (9’10″) (Recommended maximum).
Remote Sensor Cable TypePTFE (FDA food contact rated), coaxial.
Size93mm(H) x 54.5mm(W) x 8.6mm(T).
Weight41g;without External Probe.
Case MaterialPolycarbonate.
Filter
Item No. Product Description Temp. Range Unit of Measure Your Cost Purchase
ULT VFC DATALOGGER
8072-01
Low Temperature External Probe LCD Data Logger -90°C to +40°C Each
$125.09
Add to Cart This item is on back-order. Allow 5-10 business days. Please call us at 800-328-8378 with any questions about availability. In Stock